Product Code/SKU
Code
0004025400
Descriptions
Digital block heater for one aluminum block provides precise temperature control in small vessels. It is used with PCR tubes, PCR strips, Greiner tubes, microplates and cuvettes. The block heater is ideal for melting and boiling point determination, enzyme reactions, incubation and activation of cultures, immunoassays, DNA denaturation, culture media tests, coagulation tests, blood-urea-nitrogen determinations and in situ hybridizations. Stepless adjustable temperature up to 120 °C.
Scope of delivery
- Dry Block Heater 4
- Block remove tool
- PT 1000.60 Temperature sensor, stainless steel
*Please click "Request Form" for complete information regarding Prices and Stock.
Technical Specifications
| Number of blocks | 4 |
| Heat output | 412 |
| Heating temperature range | room temp. +5° - 120 °C |
| Temperature display | yes |
| Adjustment and display resolution | 1 K |
| Connection for ext. temperature sensor | CT (DIN12878) |
| PT 1000 deviation;DIN EN 60751 Kl. A | ≤ ± (0.15 +0.002xITI) K |
| Temperature stability within the blocks at 37°C * | ±0.2 °C |
| Temperature stability within the blocks at 60°C * | ±0.4 °C |
| Temperature Homogenity @ 37°C * | 0.3 K |
| Temperature Homogenity @ 60°C * | 0.6 K |
| Heatingrate / Heat up time with external sensor * | 4 K/min |
| Set-up plate material | Aluminium alloy |
| Set-up plate dimensions | 96 x 304 mm |
| Fixed safety circuit | 150 °C |
| Timer | yes |
| Time setting range | 1 - 5999 min |
| Measured in IKA DBH Testingblock | Refer Manual for more details |
| Dimensions (W x H x D) | 151 x 73 x 456 mm |
| Weight | 2 kg |
| Permissible ambient temperature | 5 - 40 °C |
| Permissible relative humidity | 80 % |
| Protection class according to DIN EN 60529 | IP 21 |
| Voltage | 220 - 240 V |
| Frequency | 50/60 Hz |
| Power input | 412 W |