Osmometry is the measurement of the osmotic strength of a solution, which refers to the concentration of solute particles in it. This is typically done using an instrument called an osmometer. Osmometry is widely used in fields like chemistry, biology, and medicine to determine properties such as molecular weight, hydration levels, or the concentration of dissolved substances in samples like blood or urine Our instruments offer the level of automation needed to ensure maximum productivity, so labs can handle any level of patient sample testing in an efficient manner. With increasing levels of staff shortages and continued challenges, this is of utmost importance.
*Please click "Request Form" for complete information regarding Prices and Stock.
Technical Specifications
| Sample volume | 20 μL |
| Test time | 90 seconds |
| Sample capacity | Up to 20 samples |
| Resolution | 1 mOsm/kg H2O |
| Range | 0 to 2000 mOsm/kg H2O |
| Accuracy2 | 0–400 mOsm: mean value ±3 mOsm/kg H2O from nominal value 400–2000 mOsm: mean value ±0.75% |
| Precision2 | (within run) 0–400 mOsm: standard deviation ≤ 3 mOsm/kg H2O 400–2000 mOsm: CV≤ 0.75% |
| Temperature effects3 | Less than 1 mOsm/kg H2O per 5°C (9°F) ambient temperature change |
| Communications | On-board printer, integrated 2D-barcode scanner, USB2.0TypeA(3), USB2.0TypeB(1), Ethernet Port(1) |
| Supported languages | Simple Chinese, Czech, Danish, English, French, German, Greek, Italian, Japanese, Korean, Portuguese, Russian, Slovak, Spanish, Swedish, Turkish |
| Storage temperature | -40°C to +45°C (-40°F to +113°F) |
| Electrical voltage | 100 to 240 VAC (50/60 Hz) |
| Power consumption | 60 Watts |
| Dimensions (D x W x H) | 37cm x 25cm x 44cm |
| Net weight | 13.2 kg (29 lbs.) |
| Shipping weight | 19.1 kg (42 lbs.) |